The ME 520-AJ is an amplifier-conditioner for strain gauge bridge sensors. It covers all sensitivity ranges from mV/V to 30 mV/V.

Using a set of jumpers, the user can configure the sensitivity range, bridge excitation voltage, output bandwidth, voltage or F65current output type, and zero offset to implement various output functions.

A calibration notch adapted to the selected sensitivity range allows for easy gain adjustment.

ME 520 NG-AJ

Description

Signal conditioner for strain gauge sensors

Bridge supply voltage: 10 V or 5 V

Adjustable sensitivity from 0.1 mV/V to 30 mV/V

Calibration switch from 0.1 to 10 mV/V (depending on range)

Gain and zero adjustments via front-panel potentiometers

Voltage output ±10 V or current output 0/4 – 20 mA

Linearity 0.01% of full scale

Adjustable bandwidth 2 kHz or 20 kHz max.

Power supply 18 to 36 Vdc

Capable of connecting 4 350-ohm sensors in parallel

Technical Specifications

Sensor input 4- or 6-wire full-bridge strain gauges
Strain gauge bridge impedance 120 < Z < 10,000 ? (for 120 ?, 5 V max)
Bridge power supply 10 V or 5 V, max. 60 mA
Sensor cable rejection 2.10?? / ?
Input sensitivity 5 ranges from 0.1 mV/V to 30 mV/V
Fixed zero offset 4 ranges from ±50% to ±100% of the calibration potentiometer value
Adjustable zero offset ±50% minimum
Calibration step From 0.1 to 10 mV/V
Calibration step accuracy 0.01% over the 1 to 3 mV/V range, 0.1% for the others
Voltage output ±10 V
Maximum output current 5 mA
Maximum output impedance 0.2 ohm
Current output 4–20 mA or 0–20 mA
Current output dynamic range 0–10 V (Load resistance: 500 ohms at 20 mA)
Linearity 0.01% of full scale
Maximum drift relative to input < 1 µV/°C
Noise relative to input < 3 µV RMS/2 kHz < 10 µV RMS/20 kHz (typical)
Common-mode rejection 100 dB
Power supply rejection 120 dB
Bandwidth 2 kHz or 20 kHz at –3 dB (typical)
Operating temperature –10 to +60°C
Storage temperature –40°C to +70°C
DIN rail enclosure H: 99 L: 17.5 D: 112 mm
Removable screw terminals 4 x 3 terminals
Enclosure weight Approximately 110 grams
Isolated main power supply 24 Vdc (18 to 36 Vdc). Current consumption: 60 mA (excluding output)
Supply voltage isolation Max. 1000 V DC for 1 min between 0 V and GND (output).

ISO Option

Supply voltage: 22 to 24 V DC 5 mA
Supply voltage isolation Max. 1000 V DC (60 sec) between 0 V and GND; 400 V peak between 0 V and Ground & GND and Ground
Variable gain Via multiplier knob: 0 to 20,000
Isolation between output and input 400 V peak between A- (input ground) and GND output