For homogeneous, electrically conductive materials
Measurement of the length of through-cracks propagating in thin metal sheets
Determination of the crack front shape inside thick test specimens (three-dimensional application possible in certain cases)
The device essentially performs three functions:
> Supplying current to the test specimen,
> Triggering the measurement,
> Amplifying and measuring the voltage across the test specimen (3 channels).
| Current intensity applied to the test specimen | Adjustable from 0 to 20 A |
| Pulse width | Adjustable from 0 to 200 ms |
| Pulse frequency | Adjustable from 0 to 150 Hz (for a 4-ms pulse) |
| Operating modes | Internal Pulse, Sector Pulse, External Pulse |
| External signal input | ±10 V |
| Gain | From 0 to 20,000 |
| Trigger threshold | 0/20 V |
| Delay times | From 0 to 200 ms, for mains or external pulses |
| Current stability | 10?³ |
| Measured voltage | 0 to ±10 V |
| Measurement Channels | 3, including 1 that uses the other 2 as inputs (A-B) and (A/B) |
| Output Current | 5 mA |
| Measurement Channel Response Time | 0.2 to 1 s |
| Variable Gain | Via multiplier knob, 0 to 20,000 |
| Supply Voltage | 220 V ± 10 V% 50 Hz |
| 4-contact male connector | JAEGER Rapid 530 754, test lead |
| 3-contact female connector | JAEGER Miniature 530 232, channels |
| Cable clamp | JAEGER 530 331 |
| Enclosure | EURONORM 3U – 81 TE |
| Length | 445.5 mm |
| Width | 132.5 mm |
| Depth | 430 mm |
Design office